Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2
90
3. In the Open dialog box, select the .xmlcon or .con file and click Open.
Verify that the sensors match those on your 19plus V2, auxiliary sensors are
assigned to the correct voltage channels, and calibration coefficients for all
sensors are up-to-date.
4. Click Save or Save As to save any changes to the .xmlcon or .con file.
Click Exit when done reviewing / modifying the configuration.
Strain gauge or Digiquartz with
temperature compensation.
and click Modify
to change
calibration
coefficients for
that sensor.
.xmlcon or .con file
for this CTD.
Open to select
.con file.
Save or Save As
to save current
.xmlcon or .con file
settings.
Channel/Sensor table reflects this
choice (0, 1, 2, 3, 4, 5, or 6). Must
agree with 19plus V2 setup for
VoltN= (N= 0, 1, 2, 3, 4, and 5);
see reply from GetCD or DS.
Voltage channel 0 in .xmlcon or
.con file corresponds to first
external voltage in data stream,
voltage channel 1 to second, etc.
cannot be removed
or changed to
another type of
sensor. All others are
optional.
navigation device used, and if
NMEA depth data and NMEA time
data were also appended.
Seasave adds current latitude,
longitude, and universal time code
to data header; appends NMEA
data to every scan; and writes
NMEA data to .nav file every time
Ctrl F7 is pressed or Add to .nav
File is clicked.
Note: Whether NMEA device was
connected to a deck unit or
directly to computer during data
acquisition in Seasave has no
effect on data file used by SBE
Data Processing, and therefore
has no effect on data processing.
• Surface PAR - Select if using with
deck unit connected to Surface
PAR sensor. Seasave appends
Surface PAR data to every scan.
Adds 2 channels to
Channel/Sensor table. Do not
increase External voltage
channels to reflect this; External
voltage channels reflects only
external voltages going directly to
19plus V2 from auxiliary sensor.
See Application Note 47.
• Scan time added - Select if
Seasave appended time (seconds
since January 1, 1970 GMT) to
each data scan.
Number of samples to average (samples at 4 Hz) in
Profiling mode. Used to calculate elapsed time, if
time is output parameter. Must agree with 19plus
setup (NAvg=); see reply from GetCD or DS.
Click a (non-shaded) sensor and click Select to pick a
different sensor for that channel. Dialog box with list of
sensors appears. Select sensors after number of
voltage channels have been specified above.
Interval between scans in Moored
mode. Used to calculate elapsed
time, if time is an output parameter.
Must agree with 19plus V2 setup
(SampleInterval=); see reply from
GetCD or DS.
SBE 38 (secondary temperature), WET Labs sensor
(up to 3 channels), up to 2 GTDs (dissolved oxygen or
nitrogen), or Optode. Must agree with 19plus
see reply from GetCD or DS. Channel/Sensor table
lists RS-232 sensors below voltage channels.
Must agree with 19plus V2 setup (MP for Profiling mode,
MM for Moored mode); see reply from GetCD or DS.