Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2
49
General Setup Commands (continued)
InitLogging Initialize logging - after all previous data
has been uploaded, initialize logging
before starting to sample again to make
entire memory available for recording.
InitLogging sets sample number
(SampleNumber=) and header and cast
number (HeaderNumber=) to 0
(sampling will start with sample 1 and
cast/header 1). If not set to 0, data will be
stored after last recorded sample. Do not
send InitLogging until all existing data
has been uploaded.
SampleNumber=x x= sample number for last sample in
memory. Typically only used to recover
data if you accidentally initialize logging
(using InitLogging) before uploading all
existing data. Do not send
SampleNumber=0 until all existing data
has been uploaded.
HeaderNumber=x x= header and cast number for last cast or
header in memory. Typically used to
recover data if you accidentally initialize
logging (using InitLogging or
SampleNumber=0) before uploading all
existing data. 19plus V2 can have a
maximum of 1000 stored headers.
Note that 19plus V2 writes a new header:
- each time logging is started in
Profiling Mode
- each time logging is started and after
every 2000 samples are stored in memory
in Moored Mode.
Biowiper=x x=Y: Configuration includes WET Labs
ECO-FL fluorometer with Bio-Wiper
(voltage or RS-232 auxiliary sensor). With
this setup, 19plus V2 is powered longer for
DS and GetSD, providing sufficient time
for Bio-Wiper to open and then shut again
if in Moored mode if Bio-Wiper is set up
to take 1 measurement for each sample
(see Application Note 72).
x=N (default): No ECO-FL with
Bio-Wiper.
QS Quit session and place 19plus V2 in
quiescent (sleep) state. Main power turned
off. Data logging and memory retention not
affected.
• The 19plus V2 requires verification
when InitLogging,
SampleNumber=, or
HeaderNumber= are sent.
19plus V2 responds:
this command will change the
scan length and/or initialize
logging. Repeat the command to
verify.
Type the command again and press
the Enter key to proceed. The
19plus V2 responds:
Scan length has changed,
initializing logging
• Do not initialize logging until all
data has been uploaded. These
commands do not delete data; they
reset the data pointer. If you
accidentally initialize logging
before uploading, recover data as
follows:
1. Set SampleNumber=a and
HeaderNumber=b, where a and b are
your estimate of number of samples
and casts in memory.
2. Upload data. If a is more than actual
number of samples or b is more than
actual number of casts in memory,
data for non-existent samples/casts
will be bad, random data. Review
uploaded data file carefully and delete
any bad data.
3. If desired, increase a and/or b and
upload data again, to see if there is
additional valid data in memory.
The 19plus V2 automatically enters
quiescent (sleep) state after 2 minutes
without receiving a command,
conserving battery energy if the user
does not send QS to put the 19plus V2
to sleep.
If an ECO-FL with Bio-Wiper is
installed and Biowiper=N, sending
GetSD or DS will open the Bio-Wiper,
but not provide enough powered time
to close it again. If you then deploy the
instrument in Moored mode (MM) with
the Bio-Wiper open and with a delayed
start time, the ECO-FL may become
fouled because the Bio-Wiper will
remain open until the first sample is
completed.