Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2
61
Polled Sampling Commands
These commands request a single sample. The 19plus V2 always stores data
for the most recent sample in its buffer. Some polled sampling commands also
store data in FLASH memory - the 19plus V2 will not execute the store data
in FLASH memory portion of those commands while logging data.
SL Output last sample from buffer (obtained
with polled sampling, or latest sample
from logging), and leave power on.
SLT Output last sample from buffer, then take
new sample and store data in buffer.
Leave power on. Data is not stored in
FLASH memory.
TS Take new sample, store data in buffer,
output data, and leave power on. Data is
not stored in FLASH memory.
TSS Take new sample, store data in buffer
and FLASH memory, output data, and
turn power off.
TSSOn Take new sample, store data in buffer
and FLASH memory, output data, and
leave power on.
GetLastSamples:x Output last x samples from FLASH
memory. If x is greater than the number of
samples in memory, 19plus V2 outputs all
samples in memory. If x is omitted,
19plus V2 outputs just the last sample.
You do not need to stop logging
(magnetic switch or Stop command)
before sending GetLastSamples:x.
Notes:
The 19plus V2 has a buffer that
stores the most recent data
samples. Unlike data in the
FLASH memory, data in the
buffer is erased upon removal or
failure of power.
Pump operation for polled
sampling is defined by
MooredPumpMode=, even if
your 19plus V2 is set to Profiling
Mode (MP). Thus, the 19plus V2
ignores the minimum conductivity
frequency (MinCondFreq=) and
pump delay (PumpDelay=) for
polled sampling.
Leave power on in SL, SLT, TS,
and TSSOn descriptions refers to
power for the 19plus V2 as well
as for the pump and any auxiliary
sensors. Power remains on until
QS is sent or the instrument
times out (after 2 minutes).