Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2
45
Status Commands (continued)
DS
Get and display operating status and setup,
which vary depending on whether in Profiling
or Moored mode.
List below includes, where applicable,
command used to modify parameter.
Profiling Mode (MP)
• Firmware version, serial number, date and time
[DateTime=]
• Voltages and currents (main and back-up
lithium battery voltages, currents – operating,
pump, external voltage sensors, RS-232 sensor)
• Logging status (not logging, logging, waiting to
start at . . ., or unknown status)
• Number of scans to average [NAvg=]
• Number of samples, sample space, and number
of casts in memory
• Profiling mode [MP], minimum conductivity
frequency for pump turn-on [MinCondFreq=],
and pump turn-on delay [PumpDelay=]
• Begin logging automatically when external
power applied [AutoRun=]?
Ignore magnetic switch position for
starting/stopping logging [IgnoreSwitch=]?
• Battery type [BatteryType=] and cut-off
voltage
• Pressure type [PType=] and range [PRange=]
• Sample RS-232 sensor - SBE 38 secondary
temperature sensor [SBE38=], WET Labs
[WetLabs=], Optode [Optode=], Gas Tension
Device [GTD=], or dual Gas Tension Devices
[DualGTD=]?
• Sample external voltages 0, 1, 2, 3, 4, and 5
[Volt0= through Volt5=]?
• Show entered commands as you type [Echo=]?
• Output format [OutputFormat=]
• Output salinity [OutputSal=] and sound
velocity [OutputSV=] with each sample?
(only if output format = converted decimal or
converted XML UVIC)
• Output sigma-t, voltage, and current with each
sample? [OutputUCSD=]
(only if output format = converted decimal or
converted XML UVIC; and if set to Y)
Example: Profiling mode (user input in bold, command used to modify parameter in parentheses)
DS
SBE 19plus V 2.3 SERIAL NO. 4000 15 Mar 2011 14:02:13 [DateTime=]
vbatt = 10.1, vlith = 8.9, ioper = 61.9 ma, ipump = 20.8 ma,
iext01 = 76.2 ma,
status = not logging
number of scans to average = 1 [NAvg=]
samples = 10, free = 4386532, casts = 1 [can clear with InitLogging]
mode = profile, minimum cond freq = 3000, pump delay = 60 sec [MP, MinCondFreq=, PumpDelay=]
autorun = no, ignore magnetic switch = no [AutoRun=, IgnoreSwitch=]
battery type = alkaline, battery cutoff = 7.5 volts [BatteryType=]
pressure sensor = strain gauge, range = 1000.0 [PType=, PRange=]
SBE 38 = no, WETLABS = no, OPTODE = no, Gas Tension Device = no [SBE38=, WetLabs=, Optode=, GTD=, DualGTD=]
Ext Volt 0 = yes, Ext Volt 1 = no [Volt0= and Volt1=]
Ext Volt 2 = no, Ext Volt 3 = no [Volt2= and Volt3=]
Ext Volt 4 = no, Ext Volt5 = no [Volt4= and Volt5=]
echo characters = yes [Echo=]
output format = converted decimal [OutputFormat=]
output salinity = no, output sound velocity = no [OutputSal=, OutputSV=]
• The DS response contains similar
information as the combined
responses from GetSD and GetCD,
but in a different format.
• Sending DS causes the pump to turn
on for a moment, so that the
19plus V2 can measure and output
the pump current. Because the
pump is designed to be water
lubricated, you will hear a noise
when the impeller spins in air.
Running the pump dry for such a
short time will not harm the pump.
• In the examples below, only voltage
channel 0 is enabled, so external
voltage current iext2345= (for
channels 2, 3, 4, and 5) is not
shown.
• In the examples below, no RS-232
sensor is enabled, so RS-232
sensor current iserial= is not shown.
• If the 19plus V2 is set up for dual
GTDs, the DS response shows:
Dual Gas Tension Device = Yes
• If the 19plus V2 is set up with a WET
Labs ECO-FL fluorometer with Bio-
Wiper (Biowiper=Y), the DS
response shows: wait
4 seconds for biowiper to
close before it measures the
enabled external voltage currents.