Status Commands (continued)
Moored Mode (MM)
• Firmware version, serial number, date
and time [DateTime=]
• Voltages and currents (main and back-up
lithium battery voltages, currents –
operating, pump, external voltage sensors,
RS-232 sensor)
• Logging status (not logging, logging, waiting
to start at . . ., or unknown status)
• Sample interval [SampleInterval=] and
number of measurements to take and average
per sample [NCycles=]
• Number of samples and available sample
space in memory
• Moored mode [MM], pump turn-on
parameter [MooredPumpMode=], and
pump turn-on delay
[DelayBeforeSampling=]
• Transmit data real-time?
[MooredTxRealTime=]
• Battery type [BatteryType=] and battery
cut-off voltage
• Pressure type [PType=] and range
[PRange=]
• Sample RS-232 sensor - SBE 38 [SBE38=],
WET Labs [WetLabs=], Optode [Optode=],
Gas Tension Device [GTD=], or dual Gas
Tension Devices [DualGTD=]?
• Sample external voltages 0, 1, 2, 3, 4, and 5
[Volt0= through Volt5=]?
• Show entered commands on screen as you
type [Echo=]?
• Output format [OutputFormat=]
• Output salinity [OutputSal=] and sound
velocity [OutputSV=] with each sample?
(only appears if output format =
converted decimal or converted XML UVIC)
• Output sigma-t, voltage, and current with
each sample? [OutputUCSD=]
(only appears if output format =
converted decimal or converted XML UVIC;
and if set to Y)
Example: Moored mode (user input in bold, command used to modify parameter in parentheses)
DS
SBE 19plus V 2.3 SERIAL NO. 4000 15 Mar 2011 14:02:13 [DateTime=]
vbatt = = 10.1, vlith = 8.9, ioper = 61.9 ma, ipump = 20.8 ma,
iext01 = 76.2 ma,
status = not logging
sample interval = 15 seconds, number of measurements per sample = 1 [SampleInterval=, NCycles=]
samples = 0, free = 1644953 [reset with InitLogging]
mode = moored, run pump for 0.5 sec, delay before sampling = 0.0 seconds
[MM, MOOREDPUMPMODE=, DELAYBEFORESAMPLING=]
transmit real-time = yes [MooredTxRealTime=]
battery type = alkaline, battery cutoff = 7.5 volts [BatteryType=]
pressure sensor = strain gauge, range = 2000.0 [PType=, PRange=]
SBE 38 = no, WETLABS = no, OPTODE = no, Gas Tension Device = no [SBE38=, Optode=, WetLabs=, GTD=, DualGTD=]
Ext Volt 0 = yes, Ext Volt 1 = no [Volt0= and Volt1=]
Ext Volt 2 = no, Ext Volt 3 = no [Volt2= and Volt3=]
Ext Volt 4 = no, Ext Volt 5 = no [Volt4= and Volt5=]
echo characters = yes [Echo=]
output format = converted decimal [OutputFormat=]
output salinity = no, output sound velocity = no [OutputSal=, OutputSV=]