Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2
81
3. The configuration information appears on the Instrument Configuration
tab. Verify that the sensors match those on your 19plus V2, and that
auxiliary sensors are assigned to the correct channels. Click Modify to
bring up a dialog box (shown below) to change the configuration and/or to
view / modify calibration coefficients.
4. Click Save or Save As to save any changes to the .xmlcon or .con file.
Click Exit when done reviewing / modifying the configuration.
Strain gauge or Digiquartz with
temperature compensation.
and click Modify
to change
calibration
coefficients for
that sensor.
.xmlcon or .con file
for this CTD.
Open to select
.con file.
Save or Save As
to save current
.xmlcon or .con file
settings.
Channel/Sensor table reflects this
choice (0, 1, 2, 3, 4, 5, or 6). Must
agree with 19plus V2 setup for
VoltN= (N= 0, 1, 2, 3, 4, and 5);
see reply from GetCD or DS.
Voltage channel 0 in .xmlcon or
.con file corresponds to first
external voltage in data stream,
voltage channel 1 to second, etc.
cannot be removed
or changed to
another type of
sensor. All others are
optional.
navigation device used, and
select whether NMEA device is
connected directly to Deck Unit or
to computer. You can also append
NMEA depth data (3 bytes) and
NMEA time data (4 bytes) after
Lat/Lon data. Seasave adds
current latitude, longitude, and
universal time code to data
header; appends NMEA data to
every scan; and writes NMEA
data to .nav file every time Ctrl F7
is pressed or Add to .nav File is
clicked.
Notes:
1. NMEA time can only be
appended if NMEA device
connected to computer.
2. NMEA depth can only be
appended if NMEA device
connected to computer or if using
19plus V2 with PN 90488 or
90545 Interface Box.
• Surface PAR - Select if using with
deck unit connected to Surface
PAR sensor. Seasave appends
Surface PAR data to every scan.
Adds 2 channels to
Channel/Sensor table. Do not
increase External voltage
channels to reflect this; External
voltage channels reflects only
external voltages going directly to
19plus V2 from auxiliary sensor.
See Application Note 47.
• Scan time added - Select to have
Seasave append time (seconds
since January 1, 1970 GMT) to
each data scan.
Number of samples to average (samples at 4 Hz) in
Profiling mode. Used to calculate elapsed time, if
time is output parameter. Must agree with 19plus
setup (NAvg=); see reply from GetCD or DS.
Click a (non-shaded) sensor and click Select to pick a
different sensor for that channel. Dialog box with list of
sensors appears. Select sensors after number of
voltage channels have been specified above.
Interval between scans in Moored
mode. Used to calculate elapsed
time, if time is an output parameter.
Must agree with 19plus V2 setup
(SampleInterval=); see reply from
GetCD or DS.
SBE 38 (secondary temperature), WET Labs
sensor (up to 3 channels), up to 2 GTDs
(dissolved oxygen or nitrogen), or Optode.
Must agree with 19plus V2 setup; see reply
from GetCD or DS. Channel/Sensor table lists
RS-232 sensors below voltage channels.
Must agree with 19plus V2 setup (MP
for Profiling mode, MM for Moored
mode); see reply from GetCD or DS.