Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2
56
Profiling Mode Setup Commands
Profiling mode setup commands following MP apply to Profiling mode
only, and have no effect on operation if the 19plus V2 is in Moored mode.
MP Set 19plus V2 to Profiling mode.
NAvg=x x= number of samples to average
(default 1, maximum 32,767). 19plus V2
samples at 4 Hz (every 0.25 seconds) and
averages NAvg samples; averaged data is
stored in FLASH memory and transmitted
real-time.
MinCondFreq=x x= minimum conductivity frequency (Hz)
to enable pump turn-on, to prevent pump
from turning on before 19plus V2 is in
water. Pump stops when conductivity
frequency drops below
MinCondFreq=.
19plus V2 Configuration Sheet lists
uncorrected (raw) frequency output at
0 conductivity. Typical value (and factory-
set default) for
MinCondFreq= for salt
water and estuarine application is:
(0 conductivity frequency + 500 Hz).
Typical value for
MinCondFreq= for fresh
water applications is:
(0 conductivity frequency + 5 Hz).
PumpDelay=x x= time (seconds) to wait after minimum
conductivity frequency (
MinCondFreq=) is
reached before turning pump on. Pump
delay time allows time for Tygon tubing
and pump to fill with water after
19plus V2 is submerged.
Pump starts
PumpDelay= seconds after
conductivity cell’s frequency output is
greater than
MinCondFreq=.
Typical value 30 - 45 seconds; range
0 600 seconds. Default 60 seconds.
AutoRun=x x=Y: Start logging automatically when
external power applied; stop logging when
external power removed. Magnetic switch
position has no effect on logging.
x=N: Wait for command when external
power applied. Default.
IgnoreSwitch=x x=Y: Do not start or stop logging based on
position of magnetic switch. Use
StartNow, StartLater, and Stop to
control logging.
x=N: Do not ignore magnetic switch
position. Logging controlled by switch
position. Default.
Note:
The 19plus V2 requires verification
when MP is sent. 19plus V2 responds:
this command will change the
scan length and/or
initialize logging. Repeat
the command to verify.
Type the command again and press
the Enter key to proceed.
The 19plus V2 responds:
Scan length has changed,
initializing logging
Note:
See Pump Operation Profiling Mode.
Note:
To start sampling immediately after
you send AutoRun=Y (if you were
performing setup on external power):
Send QS to put 19plus V2 in
quiescent (sleep) state, and then
turn external power off and then on
again. or
Send StartNow.
Example: 19plus V2 samples every 0.25 seconds. If NAvg=2, 19plus V2
averages data from 2 samples (= 1 averaged data sample per 0.5 seconds), stores
averaged data in FLASH memory, and transmits averaged data real-time.
Note:
For a 19plus V2 with optional Quartz
pressure sensor, NAvg= must be an
even number (2, 4, etc.); increasing
NAvg= reduces the measurement
noise. See Specifications in Section 2:
Description of SBE 19plus V2 to
determine resolution.
Note:
If IgnoreSwitch=N: Sending Stop only
stops logging briefly, and then logging
resumes. To stop logging, slide the
magnetic switch off or send <CTRL Z>.