Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2
55
Output Format Setup Commands
See Data Output Formats after the command descriptions for complete details
on all the formats.
OutputFormat=x x=0: Output raw frequencies and voltages
in hexadecimal. Must use this format for
acquiring and viewing real-time data in
Seasave. When using Seaterm232’s
Upload menu, Seaterm232 always uploads
data from memory in raw hex (compatible
with SBE Data Processing), regardless of
user-programmed OutputFormat=.
x=1: Output converted (engineering units)
data in hexadecimal.
x=2: Output raw frequencies and voltages
in decimal.
x=3: Output converted (engineering units)
data in decimal.
x=4: Output pressure and scan number
only, in hexadecimal. Typically used only
for interfacing with Auto Fire Module
(AFM) and SBE 32 Carousel Water
Sampler or with SBE 55 ECO Water
Sampler (for autonomous water sampling).
x=5: Output converted (engineering units)
data in decimal, in XML.
OutputSal=x x=Y: Calculate and output salinity (psu).
Only applies if OutputFormat=3 or 5.
x=N: Do not.
OutputSV=x x=Y: Calculate and output sound velocity
(m/sec), using Chen and Millero formula
(UNESCO Technical Papers in Marine
Science #44). Only applies if
OutputFormat=3 or 5.
x=N: Do not.
OutputUCSD=x x=Y: Calculate and output density sigma-t
(kg/m
3
), battery voltage, and operating
current (mA) with data polled while
logging. Voltage and current measured
after delay before sampling, but before
sampling. Operating current measured and
output only if in Moored mode. Only
applies if OutputFormat=3 or 5.
x=N: Do not.
Note:
Output format does not affect how
data is stored in FLASH memory.
Sea-Bird’s real-time data acquisition
(Seasave) and data processing
(SBE Data Processing) software
require data in raw hexadecimal
(OutputFormat=0).
Typical use of the output format
command is:
Before beginning logging:
If you will use Seasave to view
real-time data - Set output format
to raw hex.
If you will use Seaterm232 to
view real-time data - Set output
format to converted decimal for
ease in viewing real-time data.
After stopping sampling, use
Seaterm232’s Upload menu to
upload data from memory. This
automatically uploads the data in
raw hex (regardless of the
OutputFormat= setting), so the
uploaded data is compatible with
SBE Data Processing for processing
and with Seasave for viewing
archived data.