Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2
54
RS-232 Sensor Setup Commands (continued)
Setup for Aanderaa Optode:
19plus V2 must be set to 9600 baud (BaudRate=9600) if interfacing with an
Optode.
Set up Optode to interface with 19plus V2, before you connect it to
19plus V2. Connect Optode directly to computer, power with an external
power supply, and (following directions provided by Aanderaa) set:
Communication to RS-232.
Sample interval to 1 for Profiling mode, or 5 for Moored mode.
Note that the Optode samples at a maximum of 1 Hz (1 sample/sec).
In Profiling mode, 19plus V2 samples at 4 Hz; therefore, Optode output in
19plus V2 data stream will only change after every 4 scans.
Output to model number, serial number, and oxygen concentration
(micromoles/liter) in decimal format. Disable all other output (air
saturation, temperature, raw data, and text).
Connect Optode to SBE 19plus V2 Auxiliary RS-232 Input connector.
In 19plus V2, set Optode=Y to enable interface.
Optode=x x=Y: Enable RS-232 Optode.
x=N: Do not enable Optode.
SendOptode=command Command 19plus V2 to send command to
Optode and receive response; command
can be any command recognized by
Optode (see Optode manual). Reply from
Optode is limited to one line.
Example: Send stop command to Optode to interrupt sampling, and then sample
temperature command to Optode (user input in bold)
SENDOPTODE=do stop
Sending Optode: do stop
Optode RX = 4330F 139 263.160 (still sampling)
SENDOPTODE=do stop
Sending Optode: do stop
Optode RX = # (stopped sampling)
SENDOPTODE=get temperature
Sending Optode: get temperature
Optode RX = Temperature(Deg.C) 4330F 139 2.254194E+01
Notes:
When setting up the configuration
(.xmlcon or .con) file In Seasave
and/or SBE Data Processing, select
Oxygen, Optode for the Serial RS-
232C sensor. Enter the serial
number, calibration date, and
information required for salinity and
depth corrections. The internal
salinity must match the value you
programmed into the Optode (the
value is ignored if you do not enable
the Salinity correction). If you enable
Salinity correction, our software
corrects the oxygen output from the
Optode based on the actual salinity
(calculated from the CTD data). If
you enable Depth correction, our
software corrects the oxygen output
from the Optode based on the depth
(calculated from the CTD data).
You must send do stop to the
Optode to interrupt oxygen sampling
before sending commands with
SendOptode=. You may need to
send the stop command several
times.
(Moored Mode) Power turn-on
defined by DelayBeforeSampling=
does not apply to the RS-232
Aanderaa Optode, because turning
on power to the Optode causes it to
sample. The 19plus V2 turns on
power to the Optode when the
19plus V2 is ready to acquire the
Optode data, after all CTD and other
external sensor data is acquired.