Sea-Bird Electronics SBE 38 Manual de usuario Pagina 63

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 124
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 62
Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2
63
Testing Commands continued
TWetLabs Measure WET Labs RS-232 sensor. Output
varies, depending on sensor type.
ECO triple sensor:
Column
Output
1 – 2
Date and time
3
Wavelength sensor 1
4
Raw signal counts sensor 1
5
Wave length sensor 2
6
Raw signal counts sensor 2
7
Wave length sensor 3
8 Raw signal counts sensor 3
9
Thermistor counts
ECO dual sensor:
Column
Output
1 – 2
Date and time
3
Wavelength sensor 1
4
Raw signal counts sensor 1
5
Wave length sensor 2
6
Raw signal counts sensor 2
7
Thermistor counts
ECO single sensor:
Column
Output
1 – 2
Date and time
3
Wavelength sensor 1
4
Raw signal counts sensor 1
5
Thermistor counts
WETStar: Counts
C-Star:
Column
Output
1
Sensor serial number
2 Reference counts
3
Signal counts
4
Corrected signal raw counts
5
Calculated beam c, inverse meters
6
Internal thermistor, counts
TOptode Measure Aanderaa Optode, output product
number, serial number, and dissolved
oxygen (micromoles/liter).
PumpOn Turn pump on for testing purposes. Use
this command:
Before sending testing command to
obtain pumped data from sensors
plumbed with the pump, or
To test pump.
PumpOff Turn pump off for testing purposes.
Vista de pagina 62
1 2 ... 58 59 60 61 62 63 64 65 66 67 68 ... 123 124

Comentarios a estos manuales

Sin comentarios