Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2
50
Pressure and Voltage Sensor Setup Commands
PType=x Pressure sensor type (set at factory).
x=1: Strain-gauge.
x=3: Quartz with temperature
compensation. Requires even NCycles=
(2, 4, 6, etc.) for Moored mode and
even NAvg= for Profiling mode.
Volt0=x x=Y: Enable external voltage 0.
x=N: Do not.
Volt1=x x=Y: Enable external voltage 1.
x=N: Do not.
Volt2=x x=Y: Enable external voltage 2.
x=N: Do not.
Volt3=x x=Y: Enable external voltage 3.
x=N: Do not.
Volt4=x x=Y: Enable external voltage 4.
x=N: Do not.
Volt5=x x=Y: Enable external voltage 5.
x=N: Do not.
Notes:
The 19plus V2 requires verification
when these commands (PType=
through Volt5=) are sent. The
19plus V2 responds:
this command will change
the scan length and/or
initialize logging. Repeat
the command to verify.
Type the command again and press
the Enter key to proceed. The
19plus V2 responds:
Scan length has changed,
initializing logging
The 19plus V2 configuration
(.xmlcon or .con) file must match this
selection of pressure sensor and
external voltages when viewing real-
time data in Seasave or processing
uploaded data. View and edit the
configuration file in Seasave or SBE
Data Processing. These parameters
are factory-set to match the ordered
instrument configuration.
External voltage numbers 0, 1, 2, 3,
4, and 5 correspond to wiring of
sensors to a voltage channel on the
19plus V2 end cap (see Dimensions
and End Cap Connectors in Section
2: Description of SBE 19plus V2).
However, in the .xmlcon or .con file,
voltage 0 is the first external voltage
in the data stream, voltage 1 is the
second, etc.
Example: Enable voltage sensors wired to channels 0 and 3 on end cap (user
input in bold).
VOLT0=Y
VOLT1=N
VOLT2=N
VOLT3=Y
VOLT4=N
VOLT5=N
There will be 2 external sensor voltages in data stream. In .xmlcon or .con file
(in SBE Data Processing or Seasave), indicate 2 external voltage channels.
Voltage 0 corresponds to sensor wired to external voltage channel 0; voltage 1
corresponds to sensor wired to external voltage channel 3.