Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2
58
Logging Commands
Logging commands direct the SBE 19plus V2 to sample data. When
commanded to start sampling, the 19plus V2 takes samples and stores the data
in its FLASH memory. Operation is dependent on the mode and setup.
Profiling Mode (MP)
While logging, the 19plus V2 transmits real-time data, and does not enter
quiescent (sleep) state between samples. The 19plus V2 provides several
methods for starting and stopping logging, depending on the settings for
IgnoreSwitch= and AutoRun=. Logging starts approximately 1 - 2 seconds
after it is commanded.
IgnoreSwitch= AutoRun=
To Start Logging:
To Stop Logging:
N N
Slide magnetic
switch on.
Slide magnetic switch off, or
send Stop.
Y
N
Send StartNow, or
StartDateTime=
and StartLater.
Send Stop.
Y or N
Y
Turn on external
power.
Turn off external power, or
(if you want to send
commands to check or
modify setup) Send Stop.
The first time logging is started after receipt of the initialize logging command
(InitLogging), data recording starts at the beginning of memory and any
previously recorded data is written over. When logging is stopped, recording
stops. Each time logging is started again, recording continues, with new data
stored after the previously recorded data and a new header written to indicate
the incremented cast number, date, time, and sample numbers contained in the
cast. The maximum number of casts that can be taken is 1000.
Moored Mode (MM)
While logging, the 19plus V2 transmits real-time data if
MooredTxRealTime=Y. The 19plus V2 enters quiescent (sleep) state
between samples.
To start logging, use StartNow; logging starts SampleInterval= seconds after
receipt of StartNow. Alternatively, use StartDateTime= and StartLater to
start logging at a designated date and time. The first time logging is started
after receipt of the initialize logging command (InitLogging), data recording
starts at the beginning of memory and any previously recorded data is written
over. When Stop is sent, recording stops. Each time StartNow or StartLater
is sent again, recording continues, with new data stored after the previously
recorded data. A new header is written each time logging starts and every
2000 samples thereafter. A maximum of 1000 headers can be written; if the
maximum number of headers is reached but there is still room for samples in
FLASH memory, the 19plus V2 continues to sample and store sample data in
FLASH memory without writing additional headers.
Notes:
In Seaterm232, to save real-time
data to a file, click the Capture menu
before beginning logging.
If the FLASH memory is filled to
capacity, data sampling and
transmission of real-time data
continue, but excess data is not
saved in memory. The 19plus V2
will not overwrite data stored in
memory.
If the 19plus V2 is sampling and the
voltage is less than the cut-off
voltage (see BatteryType= for cut-
off values for various battery types),
the 19plus V2 halts logging and
displays WARNING: LOW BATTERY
VOLTAGE.
Note:
For Moored mode, the magnetic switch
should be left off, but it has no effect
on logging. If the switch is turned on
while the 19plus V2 is in quiescent
state, the CPU enters the awake state,
but logging does not begin. If no
communications are established, the
19plus V2 times out and enters
quiescent state after 2 minutes.
Note:
Sea-Bird ships the 19plus V2 with
AutoRun=N (it will not automatically
start sampling when external power is
applied). If you send AutoRun=Y:
Send QS to put 19plus V2 n
quiescent (sleep) state, then turn
power off and then on again to start
sampling, or
Send StartNow to start sampling.