Manual revision 006 Section 5: Routine Maintenance and Calibration SBE 19plus V2
101
Sensor Calibration
Sea-Bird sensors are calibrated by subjecting them to known physical
conditions and measuring the sensor responses. Coefficients are then
computed, which may be used with appropriate algorithms to obtain
engineering units. The conductivity, temperature, and pressure sensors on the
SBE 19plus V2 are supplied fully calibrated, with coefficients stored in
EEPROM in the 19plus V2 and printed on their respective Calibration
Certificates.
We recommend that the 19plus V2 be returned to Sea-Bird for calibration.
Conductivity Sensor
The conductivity sensor incorporates a fixed precision resistor in parallel with
the cell. When the cell is dry and in air, the sensor’s electrical circuitry outputs
a frequency representative of the fixed resistor. This frequency is recorded on
the Calibration Certificate and should remain stable (within 1 Hz) over time.
The primary mechanism for calibration drift in conductivity sensors is the
fouling of the cell by chemical or biological deposits. Fouling changes the cell
geometry, resulting in a shift in cell constant.
Accordingly, the most important determinant of long-term sensor accuracy is
the cleanliness of the cell. We recommend that the conductivity sensors be
calibrated before and after deployment, but particularly when the cell has been
exposed to contamination by oil slicks or biological material.
Temperature Sensor
The primary source of temperature sensor calibration drift is the aging of the
thermistor element. Sensor drift will usually be a few thousandths of a degree
during the first year, and less in subsequent intervals. Sensor drift is not
substantially dependent upon the environmental conditions of use, and
unlike platinum or copper elements the thermistor is insensitive to shock.
Note
After recalibration, Sea-Bird enters
the new calibration coefficients in the
19plus V2 EEPROM, and ships the
instrument back to the user with
Calibration Certificates showing the
new coefficients. We also ship a new
instrument configuration (.con) file,
which includes the new coefficients.