Manual revision 006 Section 4: Deploying and Operating SBE 19plus V2 SBE 19plus V2
44
Status Commands (continued)
GetHD Get and display hardware data, which is
fixed data describing 19plus V2:
• Device type, Serial number
• Manufacturer
• Firmware version
• Firmware date
• Command set version
• PCB serial number and assembly number
• Manufacture date
• Internal sensor types and serial numbers
• External voltage sensor types (DO,
fluorometer, etc.) and serial numbers
• External RS-232 sensor types (SBE 38,
WET Labs, etc.) and serial numbers
Example: (user input in bold, command used to modify parameter in parentheses)
gethd
<HardwareData DeviceType = 'SBE19plus' SerialNumber = '01906003'>
<Manufacturer>Sea-Bird Electronics, Inc.</Manufacturer>
<FirmwareVersion>2.3</FirmwareVersion>
<FirmwareDate>16 March 2011 08:50</FirmwareDate>
<CommandSetVersion>1.2</CommandSetVersion>
<PCBAssembly PCBSerialNum = 'not assigned' AssemblyNum = '41054F'/> [SetPcbSerialNum1=, SetPCBAssembly1=]
<PCBAssembly PCBSerialNum = 'not assigned' AssemblyNum = '41580'/> [SetPcbSerialNum2=, SetPCBAssembly2=]
<PCBAssembly PCBSerialNum = 'not assigned' AssemblyNum = '41056E'/> [SetPcbSerialNum3=, SetPCBAssembly3=]
<PCBAssembly PCBSerialNum = 'not assigned' AssemblyNum = '41059D'/> [SetPcbSerialNum4=, SetPCBAssembly4=]
<MfgDate>03 march 2011</MfgDate> [SetMfgDate=]
<InternalSensors>
<Sensor id = 'Main Temperature'>
<type>temperature0</type>
<SerialNumber>01906003</SerialNumber>
</Sensor>
<Sensor id = 'Main Conductivity'>
<type>conductivity-0</type>
<SerialNumber>01906003</SerialNumber>
</Sensor>
<Sensor id = 'Main Pressure'>
<type>strain-0</type> [PType=]
<SerialNumber>2580011</SerialNumber>
</Sensor>
</InternalSensors>
<ExternalSensors>
<Sensor id = 'volt 0'>
<type>not assigned</type> [SetVoltType0=]
<SerialNumber>not assigned</SerialNumber> [SetVoltSN0=]
</Sensor>
<Sensor id = 'volt 1'>
<type>not assigned</type> [SetVoltType1=]
<SerialNumber>not assigned</SerialNumber> [SetVoltSN1=]
</Sensor>
<Sensor id = 'volt 2'>
<type>not assigned</type> [SetVoltType2=]
<SerialNumber>not assigned</SerialNumber> [SetVoltSN2=]
</Sensor>
<Sensor id = 'volt 3'>
<type>not assigned</type> [SetVoltType3=]
<SerialNumber>not assigned</SerialNumber> [SetVoltSN3=]
</Sensor>
<Sensor id = 'volt 4'>
<type>not assigned</type> [SetVoltType4=]
<SerialNumber>not assigned</SerialNumber> [SetVoltSN4=]
</Sensor>
<Sensor id = 'volt 5'>
<type>not assigned</type> [SetVoltType5=]
<SerialNumber>hi there</SerialNumber> [SetVoltSN5=]
</Sensor>
<Sensor id = 'serial'>
<type>not assigned</type> [SetSerialType=]
<SerialNumber> not assigned </SerialNumber> [SetSerialSN=]
</Sensor>
</ExternalSensors>
</HardwareData>
External sensor types and serial
numbers can be changed in the field,
to accommodate changes in auxiliary
sensors cabled to the 19plus V2.
Other hardware data is typically not
changed by the user.
Comentarios a estos manuales