Sea-Bird Electronics SBE 38 Manual de usuario Pagina 102

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 124
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 101
Manual revision 006 Section 5: Routine Maintenance and Calibration SBE 19plus V2
102
Pressure Sensor
The SBE 19plus V2 includes a standard strain-gauge pressure sensor or
optional Quartz pressure sensor. These sensors are capable of meeting the
19plus V2 error specification with some allowance for aging and ambient-
temperature induced drift.
Pressure sensors show most of their error as a linear offset from zero. A
technique is provided below for making small corrections to the pressure sensor
calibration using the offset (POffset=) calibration coefficient term by comparing
19plus V2 pressure output to readings from a barometer.
Allow the 19plus V2 to equilibrate (with power on) in a reasonably constant
temperature environment for at least 5 hours before starting. Pressure sensors
exhibit a transient change in their output in response to changes in their
environmental temperature. Sea-Bird instruments are constructed to minimize
this by thermally decoupling the sensor from the body of the instrument.
However, there is still some residual effect; allowing the 19plus V2 to
equilibrate before starting will provide the most accurate calibration correction.
1. Place the 19plus V2 in the orientation it will have when deployed.
2. In Seaterm232:
A. Set the pressure offset to 0.0 (POffset=0).
B. Send TP to measure the 19plus V2 pressure 100 times and transmit
converted data in engineering units (decibars).
3. Compare the 19plus V2 output to the reading from a good barometer at the
same elevation as the 19plus V2 pressure sensor.
Calculate offset = barometer reading 19plus V2 reading
4. Enter the calculated offset (positive or negative) in two places:
In the 19plus V2 EEPROM, using POffset= in Seaterm232, and
In the configuration (.xmlcon or .con) file, using Seasave or SBE Data
Processing.
Offset Correction Example
Absolute pressure measured by a barometer is 1010.50 mbar. Pressure displayed from 19plus V2 is -2.5 dbars.
Convert barometer reading to dbars using the relationship: mbar * 0.01 = dbar
Barometer reading = 1010.50 mbar * 0.01 = 10.1050 dbar
The 19plus V2’s internal calculations and our processing software output gage pressure, using an assumed
value of 14.7 psi for atmospheric pressure. Convert 19plus V2 reading from gage to absolute by adding
14.7 psia to the 19plus V2 output:
-2.5 dbars + (14.7 psi * 0.689476 dbar/psia) = -2.5 + 10.13 = 7.635 dbars
Offset = 10.1050 7.635 = + 2.47 dbars
Enter offset in 19plus V2 and in .xmlcon or .con file.
For demanding applications, or where the sensor’s air ambient pressure
response has changed significantly, calibration using a dead-weight
generator is recommended. The end cap’s
7
/16-20 straight thread permits
mechanical connection to the pressure source. Use a fitting that has an O-ring
tapered seal, such as Swagelok-200-1-4ST, which conforms to MS16142 boss.
Note:
The pressure sensor is an absolute
sensor, so its raw output includes the
effect of atmospheric pressure
14.7 psi). As shown on the
Calibration Sheet, Sea-Bird’s
calibration (and resulting calibration
coefficients) is in terms of psia.
However, when outputting pressure
in engineering units, the 19plus V2
outputs pressure relative to the ocean
surface (i.e., at the surface the output
pressure is 0 decibars).
The 19plus V2 uses the following
equation to convert psia to decibars:
pressure (db) =
[pressure (psia) - 14.7] * 0.689476
Vista de pagina 101
1 2 ... 97 98 99 100 101 102 103 104 105 106 107 ... 123 124

Comentarios a estos manuales

Sin comentarios