Section 4: Setting Up MicroTSG SBE 45
30
Polled Sampling Commands
These commands are used by an external controller to request a sample from
the MicroTSG. The MicroTSG stores data for the most recent sample in its
RAM. The MicroTSG does not automatically go to sleep after executing
these commands. Do not send these commands if the MicroTSG is sampling
data at pre-programmed intervals (defined by Interval= and SingleSample=).
TS Take sample, hold converted data in RAM,
output converted data.
TSR Take sample, hold raw data (temperature and
conductivity only) in RAM, output raw data.
SLT Send last sample from RAM, output converted
data, then take new sample and hold converted
data in RAM.
TH Take sample, hold converted data in RAM.
SH Send held converted data from RAM.
Testing Commands
TT Measure temperature 100 times or until Esc
key is pressed, output converted data.
TC Measure conductivity 100 times or until Esc
key is pressed, output converted data.
TTR Measure temperature 100 times or until Esc
key is pressed, output raw data.
TCR Measure conductivity 100 times or until Esc
key is pressed, output raw data.